Hospitals with Isolated Power Systems have a need for the Portable Line Isolation Monitor (PS5-7)
When the existing line isolation monitor (LIM) fails in their Isolated Power Panel, the operating room is to be shut down from further use until the line isolation monitor is repaired or replaced, per NFPA
99. That can be costly if operations are cancelled. Since the line isolation monitor is on every hour of the day, they can fail without notice. Most hospitals do not carry a spare LIM, nor have the knowledge to replace them. With an Isolated Power Specialist Portable LIM (PS5-7), the operating room will have no down time while awaiting repairs. All that is necessary to make the operating room compliant is to plug in the PS5-7 into any outlet powered by the isolated power system where the faulty LIM exists. After the LIM is changed out, unplug the PS5-7 and put back on the shelf until needed again. Isolated Power Specialist would like to suggest one Portable LIM be specified for each new hospital. Model # PS5-7